Leak Standard Overview and Selection Guide:
Open Style Leak Standards
Description Open Style Leak Standards in crimped capillary or teflon permeation. These leak standards require gas supplied by the customer. They can be mounted into a production part to simulate a leak or permanantly plumbed in-line on a production leak test system. LACO is the only manufacturer of an open style helium permeation leak standard. These are ideal for permanent installation in production helium leak testing systems, and eliminate the common problem associated with crimped capillary leak standards plugging when the quality of the gas cannot be tightly controlled.
Listed below are a few of the many possible combinations of capillary leak standards. See our part number matrix for more information.
|
 |
|
 |
Features Fixed or adjustable leak rates
Can be built into a production part
Calibrated with most gases or mixtures
Filtered inlet and outlet
|
Applications For calibrating pressure decay leak detectors, helium leak detectors, and production leak test systems
Specifations and Ranges Available with capillary element (10-7 atmcc/sec or larger) or Teflon element (10-5 to 10-3 atmcc/sec helium only)
|
|
Additional Information
CalMaster Calibrated Leak Standards User Manual (PDF)
Calmaster Part Number Matrix (PDF)
|
| Part Number |
Description |
| CM11_._-41201GX |
Open Style Permeation Leak Standard, 10-4 |
10-4 |
NW 25 |
| CM21X-41009 |
Open Style Crimped Capillary Leak, 10-4 |
10-4 |
10-32 w/O-ring |
| CM21_._-3100C |
Open Style Crimped Capillary Leak, 10-3 |
10-3 |
1/4" NPT |