HELIUM LEAK DETECTORS FOR YOUR MAINTENANCE and service NEEDS 

LACO's TITANTEST™ M-Series Helium Mass Spectrometer Leak Detectors are configured to give you the features and accessories that you need specifically for maintenance and service applications—at an affordable price. All TITANTEST M-Series Helium Leak Detectors include a sniffer probe and a spray probe.

The dry pump configuration is ideal for semiconductor and other applications where oil-sealed rotary vane pumps are not permitted or may risk contaminating a system.

SPECIFICATIONS

ComponentsSpecifications
Minimum detectable helium leak rate (ultra fine vacuum mode)
<5 x 10-12 atm.cc/sec
Minimum detectable helium leak rate (sniffing mode)
5 x 10-8 atm.cc/sec
Maximum inlet pressure
25 mbar
Inlet helium pumping speed
(ultra fine mode)
2.5 L/sec
Roughing pump capacity
8 CFM Dry Scroll Pump
Inlet test port configuration
NW25
Detectable gases
He4, He3, and H2
Start-up time
< 3 minutes
Response time
< 0.5 seconds
Ambient operating temperature
50° F to 96° F
Operating voltage (single phase)
115/230 VAC, 50/60 Hz
Approx. weight (bench-top model)
100 lbs.*
Approx. dimensions (bench-top model)
22" L x 18" W x 12" H*
Interface
RS232 or RS 485, Analog Output

* = TITANTEST Unit Only—Cart weight or dimensions not included